GENERAL
|
Measurement
|
Oxygen
|
Configuration
|
Rack or table unit
|
Measuring range
|
4 linear measuring ranges free selectable, lowest span 1 %,
basis selection 0 to 1, 0 to 10, 0 to 25, and 0 to 100 vol% O2,
Manual, automatic or remote range control and range indication is possible
|
External range indication
|
Two line 16-sign LCD display, resolution 0.01 vol.% O2, continuous O2 indication and call off
|
Display text
|
Languages: English, French or German.
O2 transducer temperature, mA-signal, measuring range, time, date, failure/alarm information (and process pressure)
|
Response time for 90 % FSD
|
< 3 seconds at 60 Nl/hr air
|
Accuracy after calibration
|
Deviation ±1 % of 2 to 100 % span, ±2 % of 1 % span
|
Reproducibility
|
Deviation: <1 % of span
|
Influence of ambient temperature
|
No influence up to 50 °C
|
Influence of barometric pressure
|
No influence within the range 0.6 to 1.6 bar abs. with integrated pressure compensation
|
Influence of sample gas flow
|
Variation of gas flow between 0 to 60 Nl/hr air will cause a difference of < 0.1 vol.% O2
|
Sample gas inlet pressure
|
0.01 up to 0.6 bar g (PMA100 and 100L require positive pressure for adequate flow rate, no pump inside)
|
Sample gas outlet pressure
|
Recommendation: discharge freely into atmosphere, however max. 0.6 to 1.6 bar abs
|
Flow rate of sample gas
|
25 to 60 Nl/hr air
|
Temperature of sample gas
|
-10 to +50 °C, dry gas
|
O2 transducer temperature
|
Fixed at 55 °C
|
Ambient temperature
|
-10 to +50 °C
|
Storage temperature
|
-20 to +60 °C, relative humidity 0 to 90 %RH
|
Certification
|
European
|
CE marked
|
IP protection
|
IP 40 EN 60529
|
Electrical standard
|
EN 61010
|
Shipping
|
Shipping weight
|
11 Kg
|
MECHANICAL
|
Dimension
|
4 U H x 84 HP W x 350 mm D + approx 60 mm installation space
|
Enclosure
|
19 inch rack mounting with front handles, grey RAL 7032
|
Wetted materials
|
Platinum, Glass, 316Ti Stainless Steel, Epoxy resin, PTFE, PVDF, FPM
|
ELECTRICAL / UTILITIES
|
Power
|
Internal power unit 230 VAC standard, 115 VAC available ± 10%, 40 to 60 Hz. 35 VA
|
Connections
|
Mains: 3-pole chassis plug, signals: 4x sub-D plug
|
Cable length
|
2 m for mains supply
|
Sample gas connection
|
1/8 " NPT internal thread, option with tube connector DN 4/6 / PVDF available
|
Binary inputs free configurable
|
Potential free, 4 x 12 to 24 VDC, max. 20 mA or internal supply voltage 12 VDC
|
Flow alarm
|
Thermo-conductive flow sensor, mounted downstream O2 measuring cell
|
Status alarm
|
For min. flow alarm, O2 transducer temperature < 50 °C, central processing unit failure, pressure sensor: LED indication and potential free contact output, NO, max. 48 VDC, 500 mA, 15 W
and mA-output indication - example 22 mA
|
Alarm contact
|
For exceed/remain measuring range truncate calibration, external alarm, alternatively oxygen level alarms: LED indication and potential free contact output, NO, max. 48 VDC, 500 mA, 15 W
|
Outputs
|
Output signal
|
Selection: isolated 0 to 20, 2 to 20, 4 to 20, 4 to 20.5 mA for the selected range, max. load 500 Ω(i)
(interface RS232 type AK-communication protocol - bi-directional, option interface RS485 type)
|
Binary outputs
|
Drive of 3 external solenoid valves for calibration, 3 x 24 VDC, max. 400 mA (PMA 100L: 24 VDC, max. 400 mA)
|
All specifications subject to change
|