GENERAL
|
Measurement
|
Oxygen
|
Configuration
|
Portable
|
Measuring range
|
Selectable for 0 to 3, 0 to 10, 0 to 30, and 0 to 100 vol.% O2, linear. (Zero suppression measuring range 99 to 100 vol.% O2 standard, also available ranges 97/90/70 to 100 % for S version)
|
Response time for 90 % FSD
|
< 3 seconds at 60 Nl/hr air
|
Accuracy after calibration
|
Analogue = ±1 % of span
Digital = ±0.1 vol.% O2 error of precision
|
Reproducibility
|
Analogue = <1 % of span
Digital = ±0.1 vol.% O2 error
|
Influence of ambient temperature
|
Zero point ±0.02 vol.% O2 / °C;
Sensitivity ±0.1 vol.% O2 / °C
|
Influence of barometric pressure
|
The oxygen reading varies in direct proportion to changes of the barometric pressure
|
Influence of sample gas flow
|
Variation of gas flow between 0 to 60 Nl/hr air will cause a difference of <0.1 vol.% O2
|
Sample gas inlet pressure
|
0.01 up to 1 bar g, (PMA 10 required admission pressure for competent flow rate, no pump inside) option: PMA 10 with internal pump, capacity 0.9 Nl/min. without pressure
|
Sample gas outlet pressure
|
Outlet of analyser must discharge freely into atmosphere
|
Flow rate of sample gas
|
Max. 60 Nl/hr air, adjustable with needle valve on the flowmeter 7 to 70 Nl/hr
|
Temperature of sample gas
|
-10 to +40 °C, dry gas
|
Ambient temperature
|
-10 to +55 °C
|
Analyser temperature
|
According to ambient temperature, non heated version
|
Storage temperature
|
-20 to +60 °C, relative humidity 0 to 90 %RH
|
Certification
|
European
|
CE marked
|
IP protection
|
IP 41 EN 60529
|
Electrical standard
|
EN 61010
|
Shipping
|
Shipping weight
|
3 Kg
|
MECHANICAL
|
Dimension
|
150 H x 202 W x 260 D mm
|
Enclosure
|
Plastic housing, grey
|
Wetted materials
|
Platinum, Glass, Polypropylene, 316 (TI) Stainless Steel, FPM, Epoxy resin
|
ELECTRICAL / UTILITIES
|
Power
|
Internal power unit 230 VAC standard, 115 VAC available ±10%, 40 to 60 Hz. 8 VA option: rechargeable battery, recharged by the internal power unit. 10 hr battery capacity, 14 hr recharge time with instrument off.
|
Connections
|
mains: 3-pole chassis plug. signal: 3-pole plug
|
Cable length
|
2 m for mains supply
|
Sample gas connection
|
PP-hose nipple for DN 11 to 4 mm tube
|
Outputs
|
Output signal
|
0 to 1 VDC, non-isolated, load > 100 KΩ, for each selected range
|
Analogue/Digital indication
|
Analogue: selectable for each range with a scale of 0 to 30 and 0 to 100 %. Digital: 3 ½ digit 9 mm high LCD for 0 to 100 % O2 reading, selectivity 0.1 vol.% O2
|
All specifications subject to change
|